How to change the sampling rate/frequency?

The sampling frequency can be set between 12 samples per second and 50 samples per second.

Factory setting ->12 samples/s.

Xtrem scales process every 80ms (12,5 times a second).

50 samples per second equal processing every 20 ms.

Changing the AD/C sampling rate from 12 to 50 samples per second results in a higher noise of the measured value. Depending on the prevailing environmental conditions of the scale and the selected resolution, the digital filter (adjustable in the menu) may have to be increased to obtain a stable measured value.

The sampling rate can be changed in the settings under Filter Options.
The sampling rate can be changed in the settings under Filter Options.